C - 6 BeneVision N22/N19 Patient Monitor Operator’s Manual
C.3 SpO
2
/SpO
2
b Default Settings
Apnea Alarm switch On
Priority High, unadjustable
Print on Alarm Off
Apnea Delay Adult: 20 sec
Pediatric: 20 sec
Neonate: 15 sec
RR Source Auto
Resp Lead Adult: Auto
Pediatric: Auto
Neonate: II
Gain ×2
Speed 6.25 mm/s
Auto Threshold Detection On
Item Default Setting
Item Default Setting
SpO
2
/SpO
2
b Alarm switch On
High limit Adult: 100%
Pediatric: 100%
Neonate: 95%
Low limit 90%
Priority Med
Print on Alarm Off
SpO
2
/SpO
2
b
Desat
Alarm switch On
Low limit 80%
Priority High
Print on Alarm Off
SpO
2
Alarm switch On
High limit 10%
Priority Med
Print on Alarm Off
Satsecond (for Nellcor SpO
2
)Off
NIBP Simul Off
Sensitivity (for Mindray SpO
2
) Medium
Display PI On
Speed 25 mm/s