EasyManuals Logo

Mindray BeneVision N22 User Manual

Mindray BeneVision N22
472 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #424 background imageLoading...
Page #424 background image
C - 6 BeneVision N22/N19 Patient Monitor Operator’s Manual
C.3 SpO
2
/SpO
2
b Default Settings
Apnea Alarm switch On
Priority High, unadjustable
Print on Alarm Off
Apnea Delay Adult: 20 sec
Pediatric: 20 sec
Neonate: 15 sec
RR Source Auto
Resp Lead Adult: Auto
Pediatric: Auto
Neonate: II
Gain ×2
Speed 6.25 mm/s
Auto Threshold Detection On
Item Default Setting
Item Default Setting
SpO
2
/SpO
2
b Alarm switch On
High limit Adult: 100%
Pediatric: 100%
Neonate: 95%
Low limit 90%
Priority Med
Print on Alarm Off
SpO
2
/SpO
2
b
Desat
Alarm switch On
Low limit 80%
Priority High
Print on Alarm Off
SpO
2
Alarm switch On
High limit 10%
Priority Med
Print on Alarm Off
Satsecond (for Nellcor SpO
2
)Off
NIBP Simul Off
Sensitivity (for Mindray SpO
2
) Medium
Display PI On
Speed 25 mm/s

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals