BeneVision N22/N19 Patient Monitor Operator’s Manual 22 - 11
22.10 Setting the Gas Compensation
The presence of interfering gas affects the CO
2
measurement. To get the best possible measuring result, it is
needed to set the gas compensation. The configured concentration of the interfering gas should be in
accordance with its actual proportion.
For the microstream CO
2
module, gas compensations are not required.
• Make sure to use the appropriate compensations. Inappropriate compensations may cause
inaccurate measurement values and result in misdiagnosis.
For the sidestream CO
2
module, follow this procedure to set the gas compensation:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set the compensation according to the actual condition.
For the mainstream CO
2
module, follow this procedure to set the gas compensation:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set the following compensation according to the actual condition.
â–  Balance Gas
â—† Select Room Air when air predominates in the ventilation gas mixture.
â—† Select N2O when N
2
O predominates in the ventilation gas mixture.
â—† Select He when He predominates in the ventilation gas mixture.
â–  O2 Compensation
â—† Select Off when the amount of O
2
is less than 30%.
â—† Select an appropriate setting according to the amount of O
2
in the ventilation gas mixture.
â–  AG Compensation: enters the concentration of anesthetic gas present in the ventilation gas mixture. This
could compensate for the effect of AG on the readings.
22.11 Choosing a Time Interval for Peak-Picking
For microstream and mainstream CO
2
modules, you can select a time interval for picking the highest CO
2
as the
EtCO
2
and the lowest as the FiCO
2
.
To set the time interval, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set
Maximum Hold.
4. Toggle between
Single Breath, 10 s, 20 s and 30 s if microstream CO
2
module is configured; toggle
between
Single Breath, 10 s and 20 s if mainstream CO
2
module is configured.
â—† Single Breath: EtCO
2
and FiCO
2
are calculated for every breath.
â—† 10 s, 20 s, or 30 s: EtCO
2
and FiCO
2
are calculated using 10, 20 or 30 seconds of data.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals