BeneVision N22/N19 Patient Monitor Operator’s Manual 22 - 9
22.7.3 Setting the CO
2
Waveform
To set the CO
2
waveform, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set
Waveform Type, Speed and CO2 Scale of the CO
2
waveform.
22.7.4 Setting the RR Source
To set the respiration rate (RR) source, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set
RR Source.
When the current RR source does not have valid measurement, the system will automatically switch
RR Source to
Auto.
22.7.5 Entering the Standby Mode
You can set the CO
2
module to one of the following modes according to the module status:
â– Select Measure mode when you use the CO
2
modue for monitoring.
â– Select Standby mode when you does not use the CO
2
module to prolong the serviec life of the CO
2
module.
The default operating mode is
Measure. If you are not using the CO
2
module, you can proceed as follows to enter
the Standby mode:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set
Operating Mode to Standby.
22.7.6 Entering the Intubation Mode
When performing intubation during general anesthesia, you can enter the intubation mode in order to reduce
unnecessary alarms. To enter the intubation mode, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select
Intubation Mode.
For the details of the intubation mode, see
9.12 Intubation Mode.
22.8 Changing Settings for Sidestream and Microstream CO
2
Module)
22.8.1 Setting the Auto Standby
The monitor enters standby mode automatically after the configured period of time if no breath is detected since
the last detected breath. To set the auto standby, follow this procedure:
1. Select the CO
2
numeric area or waveform area to enter the CO2 menu.
2. Select the
Setup tab.
3. Set
Auto Standby.