BeneVision N22/N19 Patient Monitor Operator’s Manual 10 - 13
10.6.3 Displaying Arrhythmia Information
You can display the arrhythmia information in the numeric area. To do so, follow this procedure:
1. Access Tile Layout by either of the following ways:
◆ Select the Screen Setup quick key → select the Tile Layout tab.
◆ Select Main Menu quick key → from the Display column select Tile Layout.
2. Click the numeric area where you want to display the arrhythmia information, and then select Arrhythmia.
10.6.4 Changing Arrhythmia Settings
10.6.4.1 Changing Arrhythmia Alarm Settings
To set the arrhythmia alarm properties, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG menu.
2. Select the Arrhythmia tab→ Alarm tab.
3. Set the following alarm properties:
◆ Separately switch on or switch off each arrhythmia alarm.
◆ Select All On to switch on all arrhythmia alarms.
◆ Select All Off to switch off all arrhythmia alarms.
◆ Select Lethals Only to switch on all lethal arrhythmia alarms.
◆ Separately set the priority of each arrhythmia alarm.
◆ Separately switch on or switch off the recording for each arrhythmia alarm.
• You can switch off lethal arrhythmia alarms only when you have enabled Lethal Arrh Alarms Off. For
more information, see 10.6.4.2 Setting the Lethal Arrhythmia Alarms Switch.
• The priority of lethal arrhythmia alarms is always high. It is unconfigurable.
10.6.4.2 Setting the Lethal Arrhythmia Alarms Switch
To disable the arrhythmia alarms, follow this procedure:
1. Select the Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the Alarm tab→ select the Other tab.
3. From the Other block, set Lethal Arrh Alarms Off.
◆ Lethal Arrh Alarms Off is set to Disable by default. In this case, you cannot switch off arrhythmia
alarms.
◆ If you set Lethal Arrh Alarms Off to enable, you can switch off arrhythmia alarms from the ECG
menu. For more information, see 10.6.4.1 Changing Arrhythmia Alarm Settings.
Pauses/min Pause/min exceeds high limit
Irr. Rhythm End Irregular rhythm no longer detected for the irregular rhythm end delay time.
Afib End (for adult only)
Atrial fibrillation no longer detected for the Afib end delay time.
Arrhythmia message Description