7 - 6 BeneVision N22/N19 Patient Monitor Operator’s Manual
7.4.3 System Responses after Unpairing the TM80 and the Monitor
This section describes system responses at the monitor and at the TM80 after unpairing.
7.4.3.1 System Responses at the Monitor after Unpairing the TM80 and the Monitor
Once the TM80 is unpaired successfully with the monitor, the monitor responds as below:
In the Wireless Module menu, the Remove button on the right of the unpaired TM80 changes to Add.
The symbol disappears.
The ECG and SpO
2
labels on the monitor screen are without the-Tsuffix. The or symbol above
measurement values disappears.
7.4.3.2 System Responses at the TM80 after Unpairing the TM80 and the Monitor
Once the TM80 is unpaired successfully with the monitor, the TM80 responds as below:
If the TM80 is paired with the monitor via MPAN, the symbol will be changed to in the prompt
message area.
The TM80 allows all operations.
7.5 Pairing a BP10 with the Monitor
A BP10 can be paired with the monitor only via MPAN.
7.5.1 Pairing Procedure
To pair a BP10 with the monitor, follow this procedure:
1. Press the MPAN button on the right panel of the BP10. The Pairing... message displays in the message area
of the BP10.
2. Select the Main Menu quick key turn to the second page select Wireless Module from the Parame-
ters column at the monitor.
3. Select the desired BP10 under the Device Name column of the Wireless Module menu. If the desired BP10
is not displayed, select Update List.
4. Select Add on the right of the desired BP10.
Upon completion of pairing, the message Connection Completed appears.
5. Select the NIBP Start/Stop quick key on the monitor screen to start one NIBP measurement. Verify that the
NIBP measurement values displayed on the monitor screen and on the BP10 screen are consistent and for
the same patient.
WARNING
Make sure that the desired BP10 is paired with the monitor.
NOTE
When a BP10 is paired with a TM80, the BP10 will automatically disconnect with the TM80 once it is
paired with the monitor.
If you need to change the name of BP10, see BP10 NIBP Module Operator’s Manual (PN: 046-008269-
00).
When a BP10 is paired with the monitor, if the BP10 is out of the MPAN coverage area, it will be
disconnected from the monitor automatically. But when it is within the MPAN coverage area again, it
will be reconnected with the monitor automatically.