BeneVision N22/N19 Patient Monitor Operator’s Manual 23 - 7
23.7.4 Setting Auto Standby
The monitor enters the standby mode automatically after the configured period of time if no breath is detected
since the last detected breath. To set the auto standby, follow this procedure:
1. Select the AG numeric area or waveform area to enter the
Gas menu.
2. Select the desired gas tab .
3. Select the
Setup tab.
4. Set
Auto Standby.
23.7.5 Setting the AG Waveform
To set the AG waveform, follow this procedure:
1. Select the AG numeric area or waveform area to enter the
Gas menu.
2. Select the desired gas tab .
3. Select the
Setup tab.
4. Set the following properties of the AG waveform:
â—† Waveform Type (for CO
2
)
â—† Speed
â—† Scale
23.7.6 Setting the RR Source
To set the RR (respiration rate) source, follow this procedure:
1. Select the AG numeric area or waveform area to enter the
Gas menu.
2. Select the desired gas tab.
3. Select the
Setup tab.
4. Set
RR Source.
When the current RR source does not have valid measurement, the system will automatically switch
RR Source to
Auto.
23.7.7 Entering the Intubation Mode
When performing intubation during general anesthesia, you can enter the intubation mode in order to reduce
unnecessary alarms. To enter the intubation mode, follow this procedure:
1. Select the AG numeric area or waveform area to enter the
Gas menu.
2. Select
Intubation Mode from the bottom of the menu.
For the details of the intubation mode, see
9.12 Intubation Mode.
23.7.8 Changing the Gas Unit
To change the gas unit, follow this procedure:
1. Select the
Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the
Unit tab.
3. Set
CO2 Unit and O2 Unit.