9 - 10 BeneVision N22/N19 Patient Monitor Operator’s Manual
9.6.4 Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
To set the alarm delay time, follow this procedure:
1. Select the Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the Alarm tab → Other tab.
3. Set AlarmDelay.
The setting of Alarm Delay is not applied to the apnea alarms and the ST alarms. You can set Apnea Delay and
ST Alarm Delay separately.
9.6.4.1 Setting the Apnea Delay Time
To set the apnea delay time, follow this procedure:
1. Access the Setup page in either of the following waysin either of the following ways:
◆ Select the Alarm quick key → select the Setu tab.
◆ Select the Main Menu quick key → from the Alarm column select Setup.
2. Select Apnea Delay to set the apnea delay time.
9.6.4.2 Setting the ST Alarm Delay Time
1. Select the Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the Alarm tab → Other tab.
3. Set ST Alarm Delay.
9.6.5 Adjusting the Alarm Light Brightness
To adjust the alarm light brightness, follow this procedure:
1. Select the Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the Alarm tab → Other tab.
3. From the Alarm Light Brightness block, set Primary Display. If you are using a secondray display, set the
Secondary Display also. If you set alarm light brightness to Auto, the monitor automatically adjusts the
alarm light brightness according to the ambient light. The stronger the ambient light is, the brighter the
alarm light is.
9.6.6 Restoring the Default Alarm Settings
To reset all alarm settings to the defaults, follow this procedure:
1. Access the Setup page in either of the following WAYS:
◆ Select the Alarm quick key → select the Setup tab.
◆ Select the Main Menu quick key → from the Alarm column select Setup.
2. Select the Limits tab.
3. On the Limits page, select Defaults at the left bottom.