BeneVision N22/N19 Patient Monitor Operator’s Manual 9 - 9
AG EtCO
2
Same as the CO
2
module
FiCO
2
Same as the CO
2
module
awRR
(rmp)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
awRR+25 or
85 (whichever
is smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
FiAA/EtAA Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiO
2
/EtCO
2
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
FiN
2
O/
EtN
2
O
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
C.O. BT (°C) Adult: (BT - 1)
Pediatric: N/A
N/A Adult: (BT + 1)
Pediatric: N/A
N/A Same as the
measurement range
C.I. N/A N/A N/A N/A N/A
TFC N/A N/A N/A N/A N/A
RM RR (rpm) awRR × 0.5 or
6 (whichever
is greater)
N/A awRR × 1.5 or
30 (whichever
is smaller)
N/A Adult/pediatric: 6 to 55
Neonate: 10 to 90
PEEP
(cmH
2
O)
(PEEP - 5) N/A (PEEP + 5) N/A Same as the
measurement range
PIP
(cmH
2
O)
PIP - 10 N/A PIP + 10 N/A Same as the
measurement range
MVe (L/
min)
MVe - 2 N/A MVe + 2 N/A Same as the
measurement range
SvO
2
SvO
2
(%) SvO
2
- 5 N/A SvO
2
+ 5 N/A Same as the
measurement range
ScvO
2
(%) ScvO
2
- 5 N/A ScvO
2
+ 5 N/A Same as the
measurement range
rSO
2
rSO
2
Same as the
default alarm
limit
Manual
mode: same
as the default
alarm limit
Auto mode:
remains
unchanged
Same as the
default alarm
limit
Manual
mode: same
as the default
alarm limit
Auto mode:
remains
unchanged
15 to 95
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate