BeneVision N22/N19 Patient Monitor Operator’s Manual 4 - 3
• The settings of patient category and paced status always contain a default value, regardless of
whether the patient is admitted or not. Check if the setting is correct for your patient.
• For paced patients, you must set Paced to Yes. If it is incorrectly set to No, the monitor could mistake
a pace pulse for a QRS and fail to alarm when the ECG signal is too weak.
• For non-paced patients, you must set Paced to No.
4.3 Managing Patient Information
4.3.1 Entering the Patient Management Menu
Use any of the following methods to enter the Patient Management menu:
â– Select the patient information area at the top left corner of the screen.
â– Select the Patient Demographics quick key.
■Select the Main Menu quick key → from the Patient Management column select Demographics.
4.3.2 Editing Patient Information
Edit patient information after a patient has been admitted, or when patient information is incomplete, or when
you want to change patient information:
To edit patient information, follow this procedure:
1. Enter the Patient Management menu. For more information, see 4.3.1 Entering the Patient Management
Menu.
2. Edit patient information as required.
If you connect a barcode reader with your monitor, you can scan the patient’s barcode to enter the patient’s
medical record number (MRN).
• The monitor will reload the configuration if you changed the patient category.
4.3.3 Loading Patient Information from the ADT Server
If the monitor is connected with the Admit-Discharge-Transfer (ADT) server through the eGateway. You can load
patient information from ADT server to the monitor.
To load patient information from the ADT server, follow this procedure:
1. Enter the Find Patient menuin either of the following ways:
◆ Select the Main Menu quick key → from the Patient Management column select Find Patient.
â—† Select Find Patient from the Patient Management menu.
2. Input query criteria.
3. Select Query. Then a list pops up, including all the patients that meet the query criteria.
4. Select a patient from the patient list, and then select Import. Corresponding patient information in the
monitor will be updated.