A - 30 BeneVision N22/N19 Patient Monitor Operator’s Manual
A.14.17 RM Specifications
EtDes High (low limit + 0.2) to 18.0% 0.1%
EtDes Low 0 to (high limit - 0.2)%
FiDes High (low limit + 0.2) to 18.0%
FiDes Low 0 to (high limit - 0.2)%
Technique Diff-Pressure flow measurement technology
Flow
Measurement range Adult/pediatric:
Infant:
± (2 to 120) L/min
± (0.5 to 30) L/min
Accuracy Adult/pediatric:
Infant:
1.2 L/min or ±10% of the reading,
whichever is greater
0.5 L/min or ±10% of the reading,
whichever is greater
Resolution 0.1 L/min
Paw
Measurement range -20 to 120 cmH
2
O
Accuracy ±3%×reading
Resolution 0.1 cmH
2
O
MVe/MVi
Measurement range Adult/Pediatric:
Infant:
2 to 60 L/min
0.5 to 15 L/min
Resolution 0.01 L/min if airflow is less than 10 L/min, 0.1 L/min if airflow is equal to or
more than 10 L/min
Accuracy ±10% × reading
TVe/TVi
Measurement range Adult/Pediatric:
Infant:
100 to 1500 ml
20 to 500 ml
Resolution 1 ml
Accuracy Adult/pediatric:
Infant:
±10% or 15 ml, whichever is greater
±10% or 6 ml, whichever is greater
awRR
Measurement range 4 to 120 rpm
Resolution 1rpm
Accuracy 4 to 99 rpm
100 to 120 rpm
±1 rpm
±2 rpm