4 - 2 BeneVision N22/N19 Patient Monitor Operator’s Manual
4.1 Discharging a Patient
Before monitoring a new patient, discharge the previous patient. After the patient is discharged, all patient data,
including patient information, trend data, and physiological alarm information is be deleted from the monitor.
The technical alarms is reset, and monitor settings returns to their defaults. For more information, see 5.3
Setting Default Configuration.
After a patient is discharged, the monitor automatically admit a new patient.
WARNING
• Always discharge the previous patient before starting monitoring a new patient. Failure to do so can
lead to data being attributed to the wrong patient.
NOTE
• Discharging a patient deletes all history data from the monitor.
4.1.1 Auto Discharging a Patient after Monitor Power Off
You can let the monitor automatically discharge after the monitor has been switched off for a period of time.
To set the time period of discharging a patient, follow this procedure:
1. Select the Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the Patient Management tab→ select the Discharge tab.
3. Select a time for Auto Discharge When Power Off. The monitor will automatically discharge the patient
when it is turned off for the designated period of time. The default is Never. That is to say the monitor will
not discharge a patient no matter for how long the monitor has been switched off.
4.1.2 Manually Discharging a Patient
Manually discharge a patient using either of the following methods:
■ Select the Patient Demographics quick key → Discharge Patient.
■ Select the Main Menu quick key → from the Patient Management column select Discharge.
Select an item from the Discharge Patient menu:
â–  Selecting Ok: discharges the patient and returns to the normal monitoring screen.
â–  Selecting Standby, and then selecting Ok: discharges the patient and goes to the standby screen.
â–  Selecting Print End Case Report, and then selecting Ok: discharges the patient and prints the end case
report.
4.2 Admitting a Patient
The monitor admits a new patient in the following situations:
â–  After a patient is manually discharged, the monitor automatically admit a new patient.
â–  After being switched off for the selected time period, the monitor automatically discharge the previous
patient and admit a new patient at startup.
â–  If the monitor has not detected certain patient vital signs (ECG, SpO2, PR, RR, NIBP) for 30 minutes, you will
be prompted whether to start monitoring a new patient if any of the above vital signs are detected again.
â–  From the standby screen, select New patient.
Always inputs patient information as soon as the patient is admitted. For more information, see 4.3.2 Editing
Patient Information for details.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals