BeneVision N22/N19 Patient Monitor Operator’s Manual C - 5
C.1.4 QT Default Settings
C.1.5 Glasgow 12-lead ECG Algorithm Default Settings
C.2 Respiration Default Settings
Item Default Setting
QTc Alarm switch Off
High limit Adult: 500
Pediatric: 480
Neonate: 460
Priority Med
Print on Alarm Off
QTc Alarm switch Off
High limit 60
Priority Med
Print on Alarm Off
QT Analysis Off
QT Lead All
Item Default Setting
Filter Diagnostic
Baseline Drift Removal On
Tachy 100
Brady 50
Waveform Layout Standard
Median Complex Off
Measurements On
Interpretation On
Interpretation Summary On
Item Default Setting
RR Alarm switch On
High limit Adult: 30
Pediatric: 30
Neonate: 100
Low limit Adult: 8
Pediatric: 8
Neonate: 30
Priority Med
Print on Alarm Off