A - 6 BeneVision N22/N19 Patient Monitor Operator’s Manual
A.7 LEDs
A.8 Audio Indicator
A.9 Monitor Interface Specifications
A.9.1 Interface Specifications of the Main Unit
A.9.2 Interface Specifications of the Separate Primary Display
A.9.3 Interface Specifications of the Combined Primary Display
A.9.4 Interface Specifications of the Secondary Display
Alarm lamp 1 (three color-coded: red, yellow, and cyan)
AC power LED 1 (green)
Battery LED 1 (two color-coded: yellow and green)
Speaker Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and
multi-level tone modulation; alarm tones comply with IEC 60601-1-8.
AC power input 1
Network connector (LAN1, LAN2, LAN3) 3 standard RJ45 connectors, 100 Base-TX, IEEE 802.3
Serial bus connector (MSB) 6: 3 connectors and each connector contains 2 sub-connectors
USB connector 4, USB 2.0, for iView only
Satellite module rack (SMR) connector 3
Video output connector (VP1, VP2) 2, VP1 connects the secondary display. VP2 connects the display for iView
system.
Nurse call connector (NC) 1, standard BNC
Equipotential grounding terminal 1
Serial bus connector (MSB) 3
Serial bus hub connector (SBH) 1
Signal input connector (SIG1) 1
DC-in connector 1
Video output connector (VP1) 1
Serial bus connector (MSB) 1
Serial bus connector (MSB) 3
Serial bus hub connector (SBH) 1
Signal input connector (SIG1) 1. It should not be connected to any cable.
DC-in connector 1
Video output connector (VP1) 1