15 - 6 BeneVision N22/N19 Patient Monitor Operator’s Manual
15.8 Changing NIBP Settings
15.8.1 Setting the NIBP Alarm Properties
To set the NIBP alarm properties, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. In the
NIBP menu, select the Alarm tab.
3. Set the following alarm properties:
â—† Separately switch on or switch off the alarms for systolic pressure, diastolic pressure, and mean
pressure.
â—† Separately adjust alarm limits and alarm priorities for systolic pressure, diastolic pressure, and mean
pressure.
â—† Separately switch on or switch off the alarm recordings for systolic pressure, diastolic pressure, and
mean pressure.
15.8.2 Setting the Initial Cuff Inflation Pressure
To set initial cuff inflation pressure, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Select the
Setup tab.
3. Select
Initial Pressure, and then select the appropriate setting.
15.8.3 Setting the NIBP Interval
For auto NIBP measurement, you need to set the interval between two NIBP measurements. To set the NIBP
interval, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Select the
Setup tab.
3. Set
Interval. Selecting Manual switches to manual mode.
15.8.4 Selecting NIBP Start Mode
Start mode defines how NIBP auto mode works. To set the start mode, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Select the
Setup tab.
3. Set
Start Mode.
â—† Clock: after the first measurement, the monitor automatically synchronizes NIBP automatic
measurements with the real time clock. For example, if
Interval is set to 20 min, and you start NIBP
auto measurement at 14: 03, the next measurement will be taken at 14: 20, and then at 14:40, 15:00,
and so on.
â—† Interval: after the first measurement, the monitor automatically repeats measurements at set interval.
For example, if
Interval is set to 20 min, and you start NIBP auto measurement at 14:03, the next
measurement will be taken at 14:23, and then at 14:43, 15:03, and so on.
Stop STAT measurement and end
series
NIBP Start/Stop quick key. . Stop NIBP or Stop All button
NIBP STAT quick key .
Task By Quick Key From NIBP menu

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals