C - 4 BeneVision N22/N19 Patient Monitor Operator’s Manual
C.1.2.2 Arrhythmia Threshold Default Settings
C.1.3 ST Default Settings
Item
Default Setting
Adult Pediatric Neonate
Asystole Delay 5 sec 5 sec 5 sec
Tachy 120 bpm 160 bpm 200 bpm
Brady 50 bpm 75 bpm 100 bpm
Extreme Tachy 160 bpm 180 bpm 200 bpm
Extreme Brady 35 bpm 50 bpm 60 bpm
Multiform PVC Window 15 beats 15 beats 15 beats
PVCs/min 10 10 10
Pause/min 8 8 8
Pause Threshold 2.0 sec 2.0 sec 2.0 sec
AF/Irr Rhy End Time 2 min 2 min 2 min
V-Tach Rate 130 bpm 130 bpm 160 bpm
V Brady Rate 40 bpm 40 bpm 40 bpm
V-Tach PVCs 666
V-Brady PVCs 5 5 5
Item Default Setting
ST Alarm Mode Absolute
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-
V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb (ST
Alarm Mode set to
Absolute)
Alarm switch Off
High limit 0.2 mV
Low limit -0.2 mV
Priority Med
Print on Alarm Off
ST Single,
ST Dual (ST Alarm Mode set to
Relative)
Alarm switch Off
High limit 0.1 mV
Low limit -0.1 mV
Priority Med
Print on Alarm Off
ST Analysis Off
ST Segment Auto
Show Marker Off
ST Point J+60
Auto Adjust On
J48
ISO 80