BeneVision N22/N19 Patient Monitor Operator’s Manual 16 - 11
• Observe the Camino Operator’s Manual to make settings and to connect the monitor with the
patient.
• Because you can set the ICP alarm limits on this patient monitor, the ICP alarms settings on this
patient monitor may be different from those on the Camino device. Please pay special attention to
the alarms on the Camino.
• Only IBP module can be used for connecting the Camino. IBP connectors on other modules, such as
the MPM, PiCCO module, do not have this function.
16.8 IBP Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service
personnel.
• For the physiological and technical alarm messages, see D Alarm Messages.
(1) IBP module (2) ICP cable
(3) ICP adapter (4) ICP connector
(5) Camino cable (6) Camino monitor
Problem
Solution
Do not see IBP numeric area or waveform
area on the main screen
1. Check that the IBP is set to display in the Screen Setup menu. For more
information, see
3.11.2 Displaying Parameter Numerics and Waveforms.
2. Check that if the IBP parameter switch is enabled. If not, enable the IBP
measurement. For more information, see
3.11.1 Switching On or Off a
Parameter
.
3. Check the connection of IBP cable, IBP transducer and module.
4. Check that the stopcock is turned to the correct position.
5. Check that the IBP transducer has been zeroed. For more information, see
16.3.3 Zeroing the IBP transducer.