16 - 12 BeneVision N22/N19 Patient Monitor Operator’s Manual
Do not see systolic pressure and diastolic
pressure for P1/P2/P3/P4
Set Measure to All in the P1/P2/P3/P4 setup menu. For more information,
see
16.5.3 Setting the Pressure Type for Display.
IBP readings seem unstable 1. Make sure there are no air bubbles in the transducer systems.
2. Check that the transducer is properly fixed.
3. Zero the transducer again.
4. Replace a transducer.
Zeroing of IBP channel(s) fails. 1. Ensure that the channels are open to air. For more information, see
16.3.3 Zeroing the IBP transducer.
2. Replace a transducer.
Problem
Solution