BeneVision N22/N19 Patient Monitor Operator’s Manual 31 - 7
31.2.5.5 Printing an Event Report
You can print an event report either via a printer or via a recorder.
To print a report via a printer, follow this procedure:
1. Enter the events review page.
2. Select .
To print a report via a recorder, follow this procedure:
1. Select the
Main Menu quick key from the Alarm column select Setup.
2. Set Printing Duration.
3. Enter the events review page by either of the following ways:
Select the Review quick key select the Events tab.
Select the Main Menu quick key from the Review column select Events.
4. Select in the events review page to print the report.
31.2.6 Full Disclosure Review Page
You can review up to 48-hours’ waveform data on the full disclosure review page. You can view both the
compressed waveforms and full waveforms and numeric values.
31.2.6.1 Entering the Full Disclosure Review Page
Choose one of the following methods to enter the full disclosure review page:
Select the Review quick key select the Full Disclosure tab.
Select the Main Menu quick key from the Review column select Full Disclosure.
31.2.6.2 Selecting Waveforms
The full disclosure review page displays compressed waveforms by default. But before reviewing compressed
waveforms, you must select to store and display the desired waveforms. To store and display the desired
waveforms, follow this procedure:
1. Enter the full disclosure review page.
2. Select
Select Waveform.
3. In the
Storage column, select the desired waveforms to be stored in the monitor. In the Display
(Maximum: 3)
column, select the desired waveforms to be displayed on the review page.
The monitor is configured with a 2GB storage card by default. Therefore, the more waveforms
selected in the Storage column, the shorter the waveform storage time. The waveforms may not be
stored for 48 hours. Please exert caution when selecting waveforms.
After selecting waveforms, you can view compressed waveforms. In case of alarms, the background of
compressed waveform block at the alarm time is marked with a special color:
Red: high alarm priority
Yellow: medium alarm priority
Cyan: low alarm priority