D - 2 BeneVision N22/N19 Patient Monitor Operator’s Manual
D.1 Physiological Alarm Messages
This section lists major physiological alarms, their default priority, and the actions that can be taken when an
alarm occurs. Some of alarms may not be included.
D.1.1 General Physiological Alarm Messages
P
Note: XX represents a measurement or parameter label, such as HR, NIBP, PVCs, RR, SpO2, PR, and so on.
D.1.2 Arrhythmia Alarm Messages
Alarm messages Default priority Cause and solution
XX High High/Med, configurable XX value has risen above the high alarm limit or fallen below the low
alarm limit. Check the patient’s condition and check if the patient
category and alarm limit settings are correct.
XX Low High/Med, configurable
Apnea High The respiration signal was so weak that the monitor cannot perform
respiration analysis. Check the patient’s condition, module and patient
connections.
Alarm message Default priority
Asystole High
VFib/VTac High
Vtac High
Vent Brady High
Extreme Tachy High
Extreme Brady High
PVCs/min Med
Pause/min Med
R on T Med
Bigeminy Med
Trigeminy Med
Tachy Med
Brady Med
Multiform PVC Med
Vent. Rhythm Med
Nonsus. Vtac Med
Run PVCs Low
Pause Low
Couplet Prompt
PVC Prompt
Irr Rhythm Prompt
Pacer Not Pacing Prompt
Pacer Not Capture Prompt
Missed Beats Prompt
Vfib Prompt

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals