9 - 8 BeneVision N22/N19 Patient Monitor Operator’s Manual
IBP: ART/
Ao/UAP/
BAP/FAP/
LV/P1-P4
(Arterial
pressure)
IBP-S
(mmHg)
SYS × 0.68 +
10
(SYS - 15) or
45 (whichever
is greater)
SYS × 0.86 +
38
(SYS + 15) or
105
(whichever is
smaller)
Adult: 45 to 270
Pediatric: 45 to 185
Neonate: 35 to 115
IBP-D
(mmHg
(Dia × 0.68 +
6)
(Dia - 15) or
20 (whichever
is greater)
(Dia × 0.86 +
32)
(Dia + 15) or
80 (whichever
is smaller)
Adult: 25 to 225
Pediatric: 25to 150
Neonate: 20 to 90
IBP-M
(mmHg)
Mean × 0.68 +
8
(Mean - 15) or
35
(whichever is
greater)
Mean × 0.86 +
35
(Mean + 15)
or 95
(whichever is
smaller)
Adult: 30 to 245
Pediatric: 30 to180
Neonate: 25to 10
IBP: PA IBP-S
(mmHg)
SYS × 0.75 SYS × 0.75 SYS × 1.25 SYS × 1.25 3 to 120
IBP-D
(mmHg
Dia × 0.75 Dia × 0.75 Dia × 1.25 Dia × 1.25 3 to 120
IBP-M
(mmHg)
Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 120
IBP: CVP/
LAP/
RAP/
UVP/P1-
P4
(Venous
pressure)
IBP-M Mean × 0.75 Mean × 0.75 Mean × 1.25 Mean × 1.25 3 to 40
CO
2
EtCO
2
(mmHg)
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
0 to 32:
remains the
same
Same as the
measurement range
33 to 35: 29 33 to 35: 29 33 to 35: 41 33 to 35: 41 Same as the
measurement range
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
- 6)
36 to 45:
(etCO
2
+ 6)
36 to 45:
(etCO
2
+ 6)
Same as the
measurement range
46 to 48: 39 46 to 48: 39 46 to 48: 51 46 to 48: 51 Same as the
measurement range
>48: remains
the same
>48: remains
the same
>48: remains
the same
>48: remains
the same
Same as the
measurement range
FiCO
2
None None Same as the
default alarm
limit
Same as the
default alarm
limit
Same as the
measurement range
awRR
(rpm)
awRR × 0.5 or
6 (whichever
is greater)
(awRR - 10) or
30 (whichever
is greater)
awRR × 1.5 or
30 (whichever
is smaller)
(awRR+25) or
85 rpm
(whichever is
smaller)
Adult/pediatric: 6 to 55
Neonate: 10 to 90
Module Parameter Lower Limit Upper Limit Auto Limit Range
Adult/
pediatric
Neonate Adult/
pediatric
Neonate

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals