BeneVision N22/N19 Patient Monitor Operator’s Manual A - 21
HRavg 30 to 250 bpm 1
CVP 0 to 100 mmHg 1
MAP 0 to 500 mmHg 1
CCO-related Parameters Outputted by Vigileo Monitor
Name Range Resolution
CCO 1 to 20 L/min 0.1 L/min
CCI 0 to 20 L/min/m
2
0.1L/min/m
2
SVR 0 to 3000 DS/cm
5
0.1 kPa-s/l
SVRI 0 to 6000 DS•m
2
/cm
5
0.1 kPa-m
2
/
SV 0 to 300 ml 1 ml
SVI 0 to 200 ml/m
2
1 ml/m
2
SVV 0 to 99% 0.1%
CVP 0 to 100 mmHg 1
Alarm Limit Range Step
CCO High (Low limit + 0.1) to 25 L/min 0.1 L/min
CCO Low 0.3 to (high limit - 0.1)L/min
CCI High (Low limit + 0.1) to 15 L/min/m
2
0.1 L/min/m
2
CCI Low 0.1 to (high limit - 0.1)L/min/m
2
EDV High (Low limit + 10)to 800 ml 10 ml
EDV Low 0 to (high limit - 10)ml
EDVI High (Low limit + 10) to 400 ml/m
2
10 ml/m
2
EDVI Low 0 to (high limit - 10)ml/m
2
SVR High (Low limit + 20) to 5000 DS/cm
5
or (low limit + 2) to 500 kPa-s/l
20 DS/cm
5
or2 kPa-s/l
SVR Low 0 to (high limit - 20)DS/cm
5
or 0 to (high limit - 2)kPa-s/l
SVRI High (Low limit + 50) to 9950 DS•m
2
/cm
5
or(low limit + 5) to 995 kPa-s-m
2
/l
50 DS•m
2
/cm
5
or 5 kPa-s-m
2
/l
SVRI Low 0 to(high limit - 50)DS•m
2
/cm
5
or 0 to(high limit - 5)kPa-s-m
2
/l
SV High (Low limit + 5) to 300 ml 5 ml
SV Low 0 to (high limit - 5)ml
SVI High (Low limit + 5) to 200 ml/m
2
5 ml/m
2
SVI Low 0 to(high limit - 5)ml/m
2
RVEF High (Low limit + 5) to 100 % 5 %
RVEF Low 0 to(High limit - 5)%