C - 10 BeneVision N22/N19 Patient Monitor Operator’s Manual
IBP-M Alarm switch On
High limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
PA
Adult: 20 mmHg
Pediatric and neonate: 26 mmHg
CVP//pCVP/ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 10 mmHg
Pediatric and neonate: 4 mmHg
Low limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 70 mmHg
Pediatric: 50 mmHg
Neonate: 35 mmHg
PA
Adult: 0 mmHg
Pediatric and neonate: 12 mmHg
CVP/pCVP//ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 0 mmHg
Pediatric and neonate: 0 mmHg
Priority Med
Print on Alarm Off
CPP Alarm switch On
High limit Adult: 130 mmHg
Pediatric: 100 mmHg
Neonate: 90 mmHg
Low limit Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 30 mmHg
Priority Med
Print on Alarm Off
Measure (for P1, P2) All
Measure ((for P3, P4)) Mean only
Sensitivity Med
Speed 25 mm/sec
Scale(mmHg) CVP/pCVP/ICP/RAP/
LAP/UVP venous
pressure
0-20
Art/pArt/Ao/BAP/FAP/
LV/P1/P2 arterial
pressure
0-160
UAP/P3/P4 venous
pressure
0-80
PA 0-3 0
Item Default Setting