BeneVision N22/N19 Patient Monitor Operator’s Manual A - 25
Microstream CO
2
Module
Mainstream CO
2
Module
O
2
≤100 ±1 mmHg
N
2
O≤60
Hal ≤4
Sev ≤5
Iso ≤5
Enf ≤5
Des ≤15 ±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between 0 to
40mmHg. Inaccuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within
specification for breath rate below 15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath.
Standard Meet the standard of ISO 80601-2-55
CO2 Measurement range 0 to 99 mmHg
Accuracy* 0 to 38 mmHg:
39 to 99 mmHg:
±2 mmHg
±5% of the reading (0.08% increased in
error for every 1 mmHg if the reading is
more than 38)
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm, the accuracy is 4 mmHg or ±12% of
the reading, whichever is greater. for EtCO2 exceeding 18 mmHg. For respiration rate above 60 rpm, the above accuracy
can be achieved by using the CapnoLine H Set for Infant/Neonatal. In the presence of interfering gases, the above accuracy
is maintained to within 4%.
Resolution 1 mmHg
Sample flow rate ml/min
Initialization time 30 s (typical)
Response time 2.9 s (typical)
(The response time is the sum of the rise time and the delay time when using
a FilterLine of standard length)
Rise time: <200ms@50ml/min
awRR measurement range 0 to 150 rpm
awRR measurement accuracy 0 to 70 rpm:
71 to 120 rpm:
121 to 150 rpm:
±1 rpm
±2 rpm
±3 rpm
Data sample rate 40 Hz
Standard Meet the standard of ISO 80601-2-55
CO
2
Measurement range 0 to 150 mmHg
Accuracy 0 to 40 mmHg:
41 to 70 mmHg:
71 to 100 mmHg:
101 to 150 mmHg:
±2 mmHg
±5% of the reading
±8% of the reading
±10% of the reading