B - 4 BeneVision N22/N19 Patient Monitor Operator’s Manual
• The device is configured with a wireless network connector to receive wireless signal. Other devices
may interfere with this device even though they meet the requirements of CISPR.
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The customer or the user of the device should
assure that it is used in such an environment as described below.
Immunity test IEC60601 test level Compliance level Electromagnetic environment - guidance
Conduced RF
IEC61000-4-6
3 Vrms
150 kHz to 80 MHz
3Vrms
(ICG: 1Vrms)
Portable and mobile RF communications
equipment should be used no closer to any
part of the system, including cables, than the
recommended separation distance calculated
from the equation appropriate for the
frequency of the transmitter. Recommended
separation distances:
(ICG )
Radiated RF
IEC61000-4-3
3V/m
80MHz to 2.5GHz
3V/m Recommended separation distances:
80 MHz to 800 MHz:
800MHz - 2.5GHz:
Where, P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m)
b
.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey
a
should be less than the compliance
level in each frequency range
b
.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion band (2395.825 MHz to 2487.645
MHz) is exempt from the ESSENTIAL PERFORMANCE requirements, but remains safe.
a
: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the
ME EQUIPMENT or ME SYSTEM is used exceeds the applicable
RF compliance level above, the
ME EQUIPMENT or ME SYSTEM should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the
ME EQUIPMENT or
ME SYSTEM
.
b
: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m. For ICG, field strengths should be
less than 1V/m.