Waveform analysisMXO 4 Series
202User Manual 1335.5337.02 ─ 07
A reference waveform can be scaled, stretched and positioned in the diagram. The
"Properties" tab shows the original settings of the reference waveform, which are
stored together with the waveform data.
The following settings are displayed:
●
Time scale
●
Reference point
●
Record length
●
Position
●
Vertical scale
●
Offset
●
Position
●
Enhancement mode
●
Interpolation
●
Waveform arithmetic
Enhancement mode
The enhancement mode is taken from the source waveform of the reference. The
enhancement mode is set automatically by the instrument and depends on the sample
rate. As long as the waveform sample rate is not higher than the ADC sample rate, the
instrument works automatically in real time mode (RT). At higher waveform sample
rates, the instrument changes to interpolated time mode (IT), where interpolation adds
waveform points between the captured ADC samples by various mathematical meth-
ods, see Interpolation.
Reference waveforms