MAINTENANCE & SUPPLIES, Monthly, Clean SRV
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-25
(16) Turn ON the power. After executing auto rinse, confirm that all background values
are within the acceptable limits. If a background error occurs, perform additional
background checks as needed. If the error persists, clean the SRV again, or refer
to Chapter 8: Troubleshooting.
Table 5-5-1: Acceptable Background Limits
Parameters Acceptable Limits
WBC/BASO Channel 0.1 [x 10
3
/µL]
DIFF Channel 0.1 [x 10
3
/µL]
RBC 0.02 [x 10
6
/µL]
HGB 0.1 [g/dL]
PLT 10 [x 10
3
/µL]
(17) Run a quality control blood to make sure that the performance of the Sample Rotor
Valve is acceptable.
(18) Reset the SRV counter by referring to Section 9: Reset the Counter in this chapter.
NOTE: Make sure to reset the counter. If not reset, the message "Clean the
SRV" will remain on the LCD.
(19) Record the cleaning on the maintenance check list.