MAINTENANCE & SUPPLIES, Reset the Counter
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-71
(4) Press the [Reset] key on the right of the unit to reset the operation count to zero (0).
Manual Mode Next No. 1234567890123
Sysmex
Ready
DP:1234567890123
Sampler
XM HC GP DP SU
Service Data
Return
→←
Total Counter
Sheath Syringe
WB Pump
11213
11213
11213
SRV Counter
Piercer
Vacuum Orifice
Waste Chamber
WBC Detector
1250
0
1250
1250
430
Reset
Reset
Reset
Reset
Reset
< Counter >
Figure 5-9-4: Reset Counter Screen (Reset the Piercer Counter)
(5) Press [Return] key. The confirmation message window will appear.
Manual Mode Next No. 1234567890123
Sysmex
Read
DP:1234567890123
Sampler
XM HC GP DP SU
Service Data
→←
Total Counter
Sheath Syringe
WB Pump
11213
11213
11213
SRV Counter
Piercer
Vacuum Orifice
Waste Chamber
WBC Detector
1250
0
1250
1250
430
< Counter >
Execute settings.
Continue
CancelSet
Return
Reset
Reset
Reset
Reset
Reset
Figure 5-9-5: Reset Counters Confirmation Screen
(6) Press [Continue] to return to the Counter screen.
Press [Set] to reset the counter.
Press [Cancel] to cancel the counter reset and exit the Service Data program.