EasyManua.ls Logo

SYSMEX SF-3000 - Page 252

SYSMEX SF-3000
605 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
QUALITY CONTROL, System Settings
6-6 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(4) The password entry screen will appear. Input the password and press [ENTER].
The System Settings screen will appear as shown below, and the current Control
Method and Limit Display settings will be displayed.
QC Settings
OKCancel
L-J Control
Limit #
Limit %
X Control
Figure 6-4: System Settings Screen
(5) Press the box on the right of the desired setting keypad for each item. A check
mark "" is placed in the box when selected.
Control Method
X Control : This method performs two consecutive analyses of control
blood, and takes the mean (average) value of the two analyses
as one control data.
L-J Control : This method takes the data from a single analysis of control
blood as the control data.
Limit Display
Limit # : Limit (relative to TARGET) is calculated and displayed as an
absolute value (SD).
Limit % : Limit (relative to TARGET) is calculated and displayed as a
proportion (CV).
(6) Press [OK] to execute the new settings. The QC Setting window disappears.
Press [Cancel] to cancel the new settings. The QC Setting window disappears.
NOTE: When setting the system items, either X Control or L-J Control must
be selected, and this setting should not be changed until one lot of the
control blood has expired. Changing from X
Control to L-J Control or
vice-versa on a daily basis should not be done.

Table of Contents

Related product manuals