QUALITY CONTROL, Target/Limit, Auto Setting
Sysmex SF-3000 Operator's Manual -- Revised September 1995 6-23
(7) Press [Auto Limit] keypad.
The Select Range window will appear. When the "System Settings" is set as
Limit#, the selections will be 2SD and 3SD. When the Limit% is selected, 2CV
and 3CV will be displayed. (Refer to Section 3 in this chapter for the system
settings.)
2SD
3SD
Select Range
√
Cancel
Figure 6-19: Select Range Window
(8) Press the box on the right of the desired range specification. A check mark "√" is
placed. The window will disappear and the calculated Limit values using the
reference SD or CV values will be displayed in the Limit column.
Pressing [Cancel] closes the window without changing the setting.
FILE No.1 LOT[300070107]
MEAN
Reference
Q C
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
SD Target Limit#
7.65
4.76
13.9
40.5
85.1
29.2
34.3
45.3
15.0
217
8.3
8.9
12.2
5.24
1.41
1.00
2.46
6.39
Quit
↑
↓
→
←
ReturnSetting
Read
FD
Lot
Into.
GP
List
Auto Limit
Manual Change
Auto Target
WBC
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
NEUT#
LYMPH#
MONO#
EO#
BASO#
0.23
0.07
0.2
0.9
1.7
0.7
0.8
1.1
0.4
12
0.3
0.2
2.4
0.26
0.14
0.30
0.25
0.41
7.65
4.76
13.9
40.5
85.1
29.2
34.3
45.3
15.0
217
8.3
8.9
12.2
5.24
1.41
1.00
2.46
6.39
0.46
0.14
0.4
1.8
3.4
1.4
1.6
2.2
0.8
24
0.6
0.4
4.8
0.52
0.28
0.60
0.50
0.83
Help
Figure 6-20: Auto Limit Setting Screen