QUALITY CONTROL, QC Chart
Sysmex SF-3000 Operator's Manual -- Revised September 1995 6-37
UL
LL
UL
LL
UL
LL
UL
LL
UL
LL
UL
LL
Mean5/18
N=50
SD
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready Manual Mode Next No.1234567890123
Sampler
Q C
RBC
HGB
HCT
MCV
MCH
MCHC
3.85
14.11
38.54
89.42
28.57
29.33
AB
5/10
0.31
1.23
2.54
7.86
2.72
3.03
(CV)%
( 8.01)
(10.00)
( 6.65)
( 8.91)
( 9.41)
(10.65)
< Mean/SD >
Cursor Exclude Select
↑
↓
→
←
Return
Selected data will be excluded. O K ?
Cancel O K
QC01 LOT No. 12345678
Figure 6-34: Exclusion Confirmation Message Window
(5) Press [OK] to delete the data positioned under and between the 2 cursors from the
QC Chart.
Press [Cancel] key to cancel the program without excluding the data.
NOTE: The data on which the cursors are positioned are also excluded.
(6) Using the [←] and [→] keys, move the cursor to specify the range of data for
computation. Mean, SD and CV values calculated from the data (max.180)
between the cursors are displayed at the right of the screen.
(7) Press the [Select] key. Confirmation Window will appear as shown below. Press
[OK] to set the current Mean, SD and CV values as reference data for the
target/limit auto setting programs. Refer to Section 6.4: Auto Setting in this
chapter.
Calculated data will be set into
reference data. O K ?
Cancel
O K
Figure 6-35: Set Reference Confirmation Window