INTRODUCTION
1-20 Sysmex SF-3000 Operator's Manual -- Revised September 1995
13.3 Space Requirements
It is important to install the instrument in a suitable location. A poor location can
adversely affect its performance. Consider the following space requirements:
Select a location near a power source and close to a suitable drain.
Leave at least 0.5 m space on both sides, and at the rear and top of the system for
easy maintenance and servicing. A minimum of 0.2 m must be maintained between
the rear panel and the wall to allow for heat dissipation and tube clearance.
Install the Pneumatic Unit, reagents and printers in suitable places that will make your
work easy.
WARNING!: Install the Sampler Unit on a table or work bench. If the Sampler Unit
is installed without supporting desktop under the unit, there is a
possibility that the SF-3000 system could accidentally fall.
13.4 Installation Environment
The environmental requirements are listed below:
Operate the SF-3000 within a temperature range of 15 - 30°C (optimum temperature
25°C) and relative humidity of 30 - 85%.
To compensate for heat generated by the system, approximately 430 kcal/hour (1705
BTU/hour) is required in an air-conditioned environment. (This figure does not
include additional cooling capacity required for the optional printers.)
Avoid using the SF-3000 in areas of extreme temperature and direct sunlight.
Place the SF-3000 in a well ventilated location.
Avoid placing the SF-3000 near potentially interfering devices capable of emitting
radio frequencies such as a radio or television receiver.