TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-9
2.3 Troubleshooting Guide
1 . Pressure and Vacuum Errors
0.5 kg/cm
2
Error
Phenomena 0.5 kg/cm
2
pressure dose not reach the required range.
An analysis error may occur, and the "Check Stored Data" message also
appears. Analysis cannot be continued until the error is resolved.
Probable Cause 1) Incorrect adjustment of 0.5 kg/cm
2
pressure
2) 0.5 kg/cm
2
regulator is defective.
Corrective Action 1) Adjust 0.5 kg/cm
2
pressure.
(See
Chapter 9, Section 4: Main Unit Adjustment
.)
2) If adjustment of 0.5 kg/cm
2
pressure failed, the regulator is probably
defective. Contact your Sysmex service representative.
0.6 kg/cm
2
Error
Phenomena 0.6 kg/cm
2
pressure dose not reach the required range.
An analysis error may occur, and the "Check Stored Data" message also
appears. Analysis cannot be continued until the error is resolved.
Probable Cause 1) Incorrect adjustment of 0.6 kg/cm
2
pressure
2) 0.6 kg/cm
2
regulator is defective.
Corrective Action 1) Adjust 0.6 kg/cm
2
pressure.
(See
Chapter 9, Section 4: Main Unit Adjustment
.)
2) If adjustment of 0.6 kg/cm
2
pressure failed, the regulator is probably
defective. Contact your Sysmex service representative.