TROUBLESHOOTING
8-14 Sysmex SF-3000 Operator's Manual -- Revised September 1995
WBC Detector Thermal Sensor Discon.
Phenomena The thermal sensor monitoring the WBC detector is defective.
Analysis cannot be performed until the error is resolved.
Probable Cause The thermal sensor monitoring the WBC detector is defective or cut.
Corrective Action 1) In the auto mode analysis, reset the rack and press [Sampler], then
[Start] to resume analysis.
2) If the error persists, the thermal sensor replacement may be necessary.
Contact your Sysmex service representative.