TROUBLESHOOTING
8-24 Sysmex SF-3000 Operator's Manual -- Revised September 1995
SRV Initial Position Error
SRV Move Position Error
SRV Move Too Slow
Phenomena The SRV does not move to the specified position.
Analysis cannot be continued until the error is resolved.
When SRV Initial/Move Position Errors occur, an analysis error is reported,
and the "Check Stored Data" message also appears.
Probable Cause 1) SRV operation is faulty; not in the initial position, not in the moved
position, or takes too much time to reach the specified position.
2) SRV is dirty.
Corrective Action 1) Check to make sure that no tubing is touching any of the sampling valve
rotating part.
2) Perform SRV test operation. (Refer to
Section 3.2: Test SRV
in this
chapter.)
3) Clean the SRV. (Refer to
Chapter 5, Section 5.3: Clean SRV
.)
4) If the error persists, turn the Main Unit power OFF, then ON again.
5) If the error still persists, contact Sysmex service representative.