TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-95
Info
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Error List
Date
10/05/1995
10/05/1995
10/05/1995
10/06/1995
Time
10:15
10:20
10:25
Return
10/06/1995 10:25
9:45
Error
GP OFF Line
Replace Reagent Container PK (1)
Reagent Container PK (1) OK
Rack Move Error 1
Page
Up
Page
Down
Figure 8-38: Error Log Screen
(4) Use the [] and [] keys to move the cursor to the error message, for which you
want corrective action message.
(5) Press the [Info] keypad. The corrective action messages for the error the cursor is
positioned on will be displayed. Take the actions suggested to correct the error.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready Manual Mode Next No.1234567890123
Sampler
Error List
Date
10/05/1995
10/05/1995
10/05/1995
10/06/1995
Time
10:15
10:20
10:25
10/06/1995 10:25
9:45
Error
GP OFF Line
Replace Reagent Container PK (1)
Reagent Container PK (1) OK
Rack Move Error 1
Replace Reagent Container PK (1)
Probable Cause
1) There is insufficient reagent.
2) Float switch is defective.
3) Abnormality in the hydraulic system.
Corrective Action
1) Replace the reagent.
2) Check the float switch.
3) Check the hydraulic system.
Cancel
Page
Up
Page
Down
Info
Return
Figure 8-39: Error Information Screen
(6) Press [Return] to exit the program.