FUNCTIONAL DESCRIPTIONS, Main Unit
10-28 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(1) Valve Unit No. 93
(14) Valve Unit No. 91
(12) Valve Unit No. 89
(7) WBC Sample Asp. DP (1.0 mL)
(6) FDO Reagent DP (1.0 mL)
(5) FBA Lyse DP (1.0 mL)
(4) Dilution Sample Asp./Rinse DP (1.5 mL)
(13) Valve Unit No. 90
(9) SLS Lyse DP (0.5 mL)
(10) HGB Dilution DP (1.0 mL)
(11) RBC Dilution DP (2.0 mL)
(8) FDT Reagent DP (0.2 mL)
(3) 0.5 kg/cm2 Regulator
(2) 0.6 kg/cm2 Regulator
Figure 10-23: Main Unit – Left Interior (Diaphragm Pumps)
(1) Valve Unit No. 93 Assembly
Controls the fluid flow. This assembly includes Solenoid Valves SV46 through
SV53, and Master Valves MV47 through MV54.
(2) 0.6 kg/cm
2
Regulator
Adjusts the pressure to 0.6 kg/cm
2
.
(3) 0.5 kg/cm
2
Regulator
Adjusts the pressure to 0.5 kg/cm
2
.
(4) Dilution Sample Asp./Rinse DP (1.5 mL)
Aspirates RBC dilution sample from the Mixing Chamber through the SRV, and
also supplies diluent to the Rinse Cup for cleaning.
(5) FBA Lyse DP (1.0 mL)
Dispenses aliquot of STROMATOLYSER-FB into the WBC/BASO dilution
sample in the Reaction Chamber.
(6) FDO Reagent DP (1.0 mL)
Dispenses aliquot of STROMATOLYSER-FD(I) into the 4DIFF dilution sample in
the Reaction Chamber.