Page #468 background image
FUNCTIONAL DESCRIPTIONS
Sysmex SF-3000 Operator's Manual -- Added September 1995 10-39
Table 10-1: Reagent Information
Reagent Storage If Frozen... Open Shelf
Life
(Use within...)
WARNING!!!
CELLPACK 15 - 30°C Thaw & mix
thoroughly before
use.
60 days None
STROMATOLYSER-FD (I) 5 - 30°C REPLACE! 60 days None
STROMATOLYSER-FD (II) 5 - 30°C REPLACE! 60 days None
STROMATOLYSER-FB 2 - 30°C Thaw & mix
thoroughly before
use.
60 days None
SULFOLYSER 15 - 30°C Dissolve
precipitate by
warming reagent in
water bath at 30°C
with occasional
mixing.
60 days None
CELLCLEAN 15 - 30°C REPLACE! *8 months
after
manufacturing
Strong alkaline
detergent.
Avoid contact
with skin, eyes,
clothing.
NOTE: • Product life unopened after manufacturing is 12-18 months.
• Avoid direct sunlight during storage and use.
• If reagent displays any signs of contamination, cloudiness, instability
or color change, replace it.