SYSTEM SETUP, Units
11-46 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(4) Press the appropriate unit number to input the unit number into the unit space. The
numbers corresponding to a specific unit are displayed in the lower part of the
screen.
Units available for the different parameters are as given below.
Table 11-5: Available Reporting Units
Parameter Units Available Unit No.
WBC x10
2
/µL, x10
3
/µL, x10
9
/L 1, 2, 5
RBC x10
4
/µL, x10
6
/µL, x10
12
/L 3, 4, 6
HGB g/dL, g/L, mmol/L* 7, 8, 13
HCT L/L, %, R, (None) 9, 10, 15, 17
MCV fL 11
MCH pg, fmol*, amol* 12, 14, 16
MCHC g/dL, g/L, %, mmol/L* 7, 8, 10, 13
PLT x10
3
/µL, x10
4
/µL, x10
9
/L 2, 3, 5
RDW-SD fL 11
RDW-CV %, R, (None) 10, 15, 17
PDW fL 11
MPV fL 11
P-LCR %, R, (None) 10, 15, 17
DIFF% %, R, (None) 10, 15, 17
DIFF# x10
2
/µL, x10
3
/µL, x10
9
/L 1, 2, 5
CAUTION: When the Dutch SI units (mmol/L for HGB, fmol/amol for MCH,
and mmol/L for MCHC) are selected, the Main Unit power has to
be OFF, then ON to change the system units.
The units displayed on the screen will be changed without
power OFF and ON but those on the analysis results display
and output will not change.
(5) Press [QUIT] key to exit the program. The confirmation message will appear.
Press the appropriate key [Continue], [Set] or [Cancel].
Continue : To return to Units setting screen so that setting can be continued.
Set : To save the new settings. Program ends.
Cancel : To cancel the new settings. Program ends.