SYSTEM SETUP
11-58 Sysmex SF-3000 Operator's Manual -- Revised March 1996
(3) Insert the floppy disk into the drive.
(4) Press [Enter] to backup setting data onto the disk. While backup is in progress, the
message "Backup..." will appear in the window.
Auto
Manage.
Data
Criteria
Config
Print
Settings
Condition
Setting
Flag
System
Return
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Backup
Set Value
Backup Process in Progress.
Please wait . . .
Figure 11-48: FD Writing Window
(5) If the settings cannot be written onto the floppy disk, the error message shown
below will appear in the FD backup window. Insert the correct type of floppy disk
and press [Enter].
Auto
Manage.
Data
Criteria
Config
Print
Settings
Condition
Setting
Flag
System
Return
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Backup
Set Value
Cancel OK
FD write Error!
Insert FD and Press OK key.
Figure 11-49: FD Write Error Message Window
(6) Press [Cancel] to cancel processing and exit the program.
NOTE: Only the output of the setting values is possible. To restore the
setting values, contact your Sysmex service representative.