SAMPLE PROCESSING, Manual Mode Procedures
Sysmex SF-3000 Operator's Manual -- Revised September 1995 2-19
2 . Select Manual Mode
If "Manual Mode" does not appear on the LCD display, follow the procedure below to
select Manual mode.
(1) Press the Next No. keypad in the top line of the LCD. The Sample No. Setting
screen shown below will appear.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Closed Mode Next No.1234567890123
Sampler
Manual Sample Number
CANCEL
ENTER
C
3
9
6
0
1
7
4
2
8
5
Ð
QC
1234567890123
Analysis Mode
Manual Mode
Capillary Mode
Closed Mode
Figure 2-11: Sample No. Setting Screen (Manual Mode)
(2) The selected analysis mode is backlit in light-blue. Press [Manual Mode] to select
Manual mode.
3 . Input Sample ID Number
Using the Sample No. Setting Screen, input the next sample ID number to be analyzed as
described below, if needed.
(1) The next sample number to the previously analyzed sample in Manual mode
appears above the numerical keys. Clear the number by pressing [C] key and input
the sample ID number with the numerical keys, if needed.
NOTE: When the next sample No. is not set, or after the power ON, the next
sample No. is automatically set to No. 1, increasing by 1 for each
succeeding sample (e.g.
, 1, 2, 3, ...).
(2) Press [ENTER]. The next sample No. is set and the Sample No. Setting screen
will disappear. The input sample No. will be displayed in the Next No. column in
the top line of the LCD.