SAMPLE PROCESSING, Manual CP Mode Procedures
2-44 Sysmex SF-3000 Operator's Manual -- Revised September 1995
CAUTION: • If VENOJECT® II collection tubes are to be used, some engineering
change is required. Contact a Sysmex service representative.
• Recapped VENOJECT® II collection tube cannot be used. While
inverting or piercing such a collection tube, the cap may be
removed.
2 . Select Closed Mode
If "Closed Mode" does not appear on the LCD display, follow the procedure below to
select Manual mode.
(1) Press the Next No. keypad in the top line of the LCD. The Sample No. Setting
screen will appear.
(2) The selected analysis mode is backlit in light-blue. Press [Closed Mode] to select
Manual CP mode.
The messages "Not Ready" and "Mode Change" appear in the top line of the LCD.
After the system condition becomes ready for analysis, "Ready" and "Closed
Mode" will be displayed.
Sysmex
DP:1234567890123
XM HC GP DP CP
Ready
Closed Mode Next No.1234567890123
Sampler
Closed Sample Number
CANCEL
ENTER
C
3
9
6
0
1
7
4
2
8
5
Ð
QC
1234567890123
Analysis Mode
Manual Mode
Capillary Mode
Closed Mode
Figure 2-35: Sample No. Setting Screen (Manual CP/Closed Mode)