380 Keysight M8000 Series of BER Test Solutions User Guide
7 Setting up Patterns
Sequence Configuration
The sequence configuration section provides the following options:
• Sequence: Allows you to apply a sequence configuration on either
Generator, Analyzer or user-created sequences.
• Name: Provide the sequence name.
• Locations: Specify the locations for the sequence. Use the button
available to select the locations.
• Replicate: Select the replicate option (Serialized, Copy or Copy plus
Phase Adjust).
• Description: Add a description to the sequence.
Make sure not to use the unused location which are not being used in the
test setup. However, if it is used, ensure that the 'Re-Sync' parameter of
these locations is set to manual to reduce overall system load caused by
unnecessary pattern re-sync attempts.