380 Keysight M8000 Series of BER Test Solutions User Guide
7 Setting up Patterns
Sequence Configuration
The sequence configuration section provides the following options:
Sequence: Allows you to apply a sequence configuration on either
Generator, Analyzer or user-created sequences.
Name: Provide the sequence name.
Locations: Specify the locations for the sequence. Use the button
available to select the locations.
Replicate: Select the replicate option (Serialized, Copy or Copy plus
Phase Adjust).
Description: Add a description to the sequence.
NOTE
Make sure not to use the unused location which are not being used in the
test setup. However, if it is used, ensure that the 'Re-Sync' parameter of
these locations is set to manual to reduce overall system load caused by
unnecessary pattern re-sync attempts.