388 Keysight M8000 Series of BER Test Solutions User Guide
7 Setting up Patterns
Sharing Sequences
The M8070A software currently does not allow the functionality to share
the sequences. However, there is a workaround to share sequences.
Follow the given steps:
1Go to File > and click Save Instrument State... Once you save the
instrument state, the sequence settings are also saved in the Settings
folder which is created at the following path:
“Documents\Agilent\M8070A\Workspaces\Default\User\Settings”
2Share Settings folder.
3Copy the Settings folder at the same path on another system.
4Go to File > and click Recall Instrument State.... The sequences will
now appear in the Sequence Editor.