Remote control commands
R&S
®
RTE
1723User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:SENT:STYPe <SearchName>,<SMFieldType>
SEARch:TRIGger:SENT:STYPe? <SearchName>
Selects the serial message sequence to be searched for.
To enable the search for one of the serial message types, use BUS<m>:SENT:
SFORmat and enable with SEARch:TRIGger:SENT:SMSG.
Parameters:
<SMFieldType> ID | IDDT
ID
Searches for the identifier.
To set the identifier condition for the serial message, use
SEARch:TRIGger:SENT:SICN, SEARch:TRIGger:SENT:
SIMN and SEARch:TRIGger:SENT:SIMX.
IDDT
Searches for the combination of identifier and data.
To set the identifier condition, use the commands shown above.
To set the data condition, use SEARch:TRIGger:SENT:SDCN,
SEARch:TRIGger:SENT:SDMN and SEARch:TRIGger:SENT:
SDMX.
*RST:
ID
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:SIDType <SearchName>,<SMIDType>
SEARch:TRIGger:SENT:SIDType? <SearchName>
Sets the message ID format (4 bit or 8 bit) of the enhanced serial message.
Parameters:
<SMIDType> B4 | B8
*RST: B4
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SENT:SICN <SearchName>,<SMIDOperator>
SEARch:TRIGger:SENT:SICN? <SearchName>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<SMIDOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater than,
Greater or equal than. These conditions require one data pattern
to be set with SEARch:TRIGger:SENT:SIMN.