RESULT INTERPRETATION/REPORTS, IP Messages
4-4 Sysmex SF-3000 Operator's Manual -- Revised February 1997
The IP messages used by the SF-3000 are listed on the following tables.
NOTE: A [C] following the item number means that the message will appear
with samples analyzed in the Capillary Mode.
Item No. refers to the code displayed on the IP message screen in Stored Data (See
Section 3.6 in this chapter).
Table 4-2: WBC IP Messages
WBC IP MESSAGES
"Abnormal" messages
Item No. Message Abbreviation Meaning Category
1 WBC Abn Scattergram WBC Abn Scg WBC abnormal scattergram Morph.
2 Neutropenia
Neutro
Low neutrophil count Diff.
3 Neutrophilia
Neutro
High neutrophil count Diff.
4 Lymphopenia
Lympho
Low lymphocyte count Diff.
5 Lymphocytosis
Lympho
High lymphocyte count Diff.
6 Monocytosis
Mono
High monocyte count Diff.
7 Eosinophilia
Eo
High eosinophil count Diff.
8 Basophilia
Baso
High basophil count Diff.
9 [C] Leukocytopenia
Leuko
Low leukocyte count Count
A [C] Leukocytosis
Leuko
High leukocyte count Count
"Suspect" messages
Item No. Message Abbreviation Meaning Category
1 Blasts/A-Lymph? Blasts/A-Ly? Possibility that blasts, atypical
lymphocytes, or abnormal lymphocytes
are present
Morph.
2 Immature Gran? Imm Gran? Possibility that immature granulocytes
are present
Morph.
3 Left Shift? Left Shift? Possibility of left shift Morph.
4 Aty/Abn Lympho? Aty/Abn Ly? Possibility that atypical lymphocytes or
blasts are present
Morph.
6 NRBC/PLT Clumps? NRBC/PLTC? Possibility that nucleated RBC or PLT
clumps are present
Morph.
Count
Sysmex SF-3000 Operator's Manual -- Revised October 1995