RESULT INTERPRETATION/REPORTS
4-22 Sysmex SF-3000 Operator's Manual -- Revised September 1995
3) Sample No.
The sample No. is displayed below the screen title. Sample numbers can consist of up
to 13 digits, including hyphens. The following 3 types of indications appear in front of
the number.
ID : Indicates that the number was read by ID bar code reader. If a
number cannot be read by the bar code reader, "ID" is backlit in
red.
C (backlit in green) : Indicates sample analyzed in the Capillary mode.
P (backlit in green) : Indicates sample analyzed in the Closed mode or Manual CP
mode.
4) Date and Time
Date and time that the analysis results were received.
5) R: Rack No. – Tube Position Number
Rack No. and the tube position No. of the sample which was analyzed in Auto mode.
"R:0000-00" is displayed for the sample analyzed in other modes.
6) DIFF Scattergram
Two-dimensional display of DIFF scattergram, according to the DIFF Channel data by
flow cytometry using a semi-conductor laser. The horizontal and vertical axes are
marked with scales consisting of 5 equal divisions. Refer to Chapter 10, Section 6:
WBC Scattergrams.
7) WBC/BASO Scattergram
Two-dimensional display of WBC/BASO scattergram, according to the BASO Channel
data by flow cytometry using a semi-conductor laser. The horizontal and vertical axes
are marked with scales consisting of 5 equal divisions. Refer to Chapter 10, Section 6:
WBC Scattergrams.
8) RBC Particle Size Distribution
Relative height curve showing RBC distribution.
9) PLT Particle Size Distribution
Relative height curve showing PLT distribution.