MAINTENANCE & SUPPLIES, Weekly, Clean WBC Detector
5-12 Sysmex SF-3000 Operator's Manual -- Revised March 1996
(6) Open the front cover and set the stop bar in position.
WARNING! When working with the front cover open, make sure that the stop bar
is in place.
WARNING! CELLCLEAN (or 5% filtered sodium hypochlorite) is a strong
alkaline solution. Avoid contact with it. If you get it on your skin or
clothing, immediately wash it off with excess water.
(7) Use the provided syringe (Mini - Pet No. 10) to pour 1 mL each of CELLCLEAN
[or 5% filtered sodium hypochlorite (bleach)] into the reaction chambers as shown
in Figure 5-4-8.
CAUTION: Do not pour CELLCLEAN more than 2.0 mL into the chamber to
avoid overflow; this can result in system malfunction.
If reagent spills on painted surfaces of the instrument, immediately
wipe it up with a damp tissue.
Reaction Chamber 2
Reaction Chamber 1
Figure 5-4-8: Place CELLCLEAN
Sysmex SF-3000 Operator's Manual -- Revised July 1996