MAINTENANCE & SUPPLIES, Weekly, Clean WBC Detector
Sysmex SF-3000 Operator's Manual -- Revised March 1996 5-13
(8) Press the Start Switch. The cleaning sequence will start. During the cleaning
sequence, the elapsed time is displayed as shown below.
Sysmex
Running
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Cleaning WBC Detector.
0 min. / 5 min.
Maintenance
Clean
RBC Transducer
Return
Figure 5-4-9: Cleaning WBC Detector Screen (3)
(9) After the cleaning sequence is completed, the instrument automatically performs
auto rinse and returns to Ready status.
(10) Release the stop bar and close the front cover.
(11) Reset the counter of the WBC Detector. Refer to Section 9: Reset the Counter in
this chapter.
NOTE: Make sure to reset the counter. If not, the message "Clean the WBC
Detector" will remain on the LCD.
(12) Record the cleaning on the maintenance check list.