MAINTENANCE & SUPPLIES, Monthly, Clean SRV
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-23
CAUTION: When a fixed valve is removed, reagent might leak from the tubes.
Prepare on hand a cloth or tissue to absorb any reagent.
(9) Remove the front fixed valve by gently pulling toward you, then remove the
rotation valve. See Figure 5-5-15.
NOTE: If the sections of the valve are stuck together, you can remove them
by slightly pulling toward you, gently turning, and then sliding them
out.
Rotor Valve
Figure 5-5-15: Remove Rotor Valve
WARNING! CELLCLEAN (or 5% filtered sodium hypochlorite) is a strong
alkaline solution. Avoid contact with it. If you get it on your skin or
clothing, immediately wash it off with excess of water.
CAUTION: Do not use solvents other than CELLCLEAN (or 5% filtered
sodium hypochlorite (bleach)).
The sampling valves are adequately resistant to corrosion from
CELLCLEAN; however, completely wipe off the CELLCLEAN to
prevent problems in other parts of the instrument.
Do not damage the sampling valve surface in any way; this can
cause leaks.