MAINTENANCE & SUPPLIES, As Needed, RBC Transducer Cleaning Sequence
5-36 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(8) Use the provided syringe (Mini - Pet No. 10) to pour 1.0 mL of CELLCLEAN (or
5% filtered sodium hypochlorite (bleach)) into the transducer chamber as shown in
Figure 5-7-12.
CAUTION: • To avoid overflow, do not pour more than 2.0 mL CELLCLEAN into
the chamber. This can result in system malfunction.
• If reagent spills on painted surfaces of the instrument, immediately
wipe it out with a damp tissue.
Figure 5-7-12: Place CELLCLEAN
(9) Replace the transducer cap.
(10) Press the Start Switch to execute the clean transducer sequence. The cleaning
sequence will start. During the rinse sequence, the elapsed time is displayed as
shown below.
Sysmex
Running
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
<Service Sequence>
Cleaning RBC Transducer.
0 min. / 2 min.
Maintenance
Clean
RBC Transducer
Return
Figure 5-7-13: Cleaning Transducer Message Window (3)
Sysmex SF-3000 Operator's Manual -- Revised July 1996